Blood compatibility of a new type of EPC capture coronary stent coating in vitro

LI Qian,CAI Dan,DIAO Hong-ying,LIU Jian-gen,ZHANG Ji-chang,LU Yang,LI Ying,SONG Chun-li
DOI: https://doi.org/10.3877/cma.j.issn.1674-0785.2013.18.042
2013-01-01
Abstract:Objective To investigate the blood compatibility of PEG-PLA-PGL/RGD on 316L stainless steel surface in vitro so as to provide further experimental reference for its clinical application. Methods The blood compatibility was evaluated by hemolysis assay, platelet adhesion experiments, anticoagulant time measurement and protein adsorption test. Results The hemolysis rate of each group was less than 5%, which suggested they were all avirulent; APTT remarkable prolonged compared with plasma sample, showed anti-coagulant property of steel in three groups(P=0.047, 0.044, 0.036). Compared with bare steel, PEG-PLA-PGL steel had no apparent differences in any tests, meanwhile, PEG-PLA-PGL/RGD steel adsorbed less platelets but larger amount of albumin(P=0.046, 0.043). Conclusion Blood compatibility of PEG-PLA-PGL/RGD steel has been improved compared with PEG-PLA-PGL steel and bare steel. PEG-PLA-PGL/RGD might be a potential appropriate material for coronary stent.