In Vivo Study of Biocompatibility of PEG-PLA-PGL As Biodegradable Carrier of Drug Eluting Stent

刘斌,张基昌,陈宏勃,宋显晶
DOI: https://doi.org/10.3969/j.issn.1007-4287.2006.09.010
2006-01-01
Abstract:Objective To investigate the biocompatibility and biological safety of PEG-PLA-PGL polymer in vivo so as to provide experimental reference for clinical application.Methods Acute systemic toxicity test,intracutaneous stimulation test,hemolysis test,cytotoxicity test,pyrogen test,sensitivity test and muscle embedding test were performed in order to evaluating the biocompatibility and biological safety of PEG-PLA-PGL polymer.Results There was no potential allergic substance in PEG-PLAPGL polymer,and no hemolytic reaction,acute systemic toxicity effect or pyrogen reaction was observed.Mild inflammatory reaction was seen in the surrounding tissues in the early implantation stage,which completely disappeared after 12 weeks.Macrophage aggregation was not noted.The polymer was almost degraded after 16 weeks.Conclusion PEG-PLA-PGL has fine biocompatibility,which is feasible and safe for clinical reconstruction of material as carrier of coronary drug eluting stent.