Correlation between interface performance and blood compatibility of heparin-loaded poly (lactic-co-glycolic acid) coating

Wu YF,Yingying Huang,Min Qi,Zhiyong Wei,Dazhi Yang
DOI: https://doi.org/10.3321/j.issn:1673-8225.2007.09.017
2007-01-01
Abstract:Aim: To evaluate the blood compatibility of poly (lactic-co-glycolic acid) (PLGA) coatings. Methods: The experiment was conducted in the School of Materials and Engineering, Dalian University of Technology from May to June 2006. PLGA coatings and the heparin coatings on 316L stainless steel plates were prepared by dipping method. The blood compatibility was primarily evaluated by platelet adhesion test. Contact angle of distilled water and glycerin on surfaces of the samples were tested to calculate the surface energy parameters such as surface tension, interface tension and adhesion power, respectively. Additionally, the effects of these parameters on the blood compatibility were analyzed. Results: 1 The contact angles of PLGA coating, 2% heparin coating and 5% heparin coating were 78.5°, 81.5° and 83°, respectively. Surface tension, interface tension and adhesion power of heparin coatings were decreased compared with PLGA coating. 2 The adhesion platelets of PLGA coating were more, and some platelets deformed and the parapodium stretched out; the adhesion platelets of heparin coating were less without deformity or aggregation, and the adhesion platelets were decreased with the increase of heparin coating. Conclusion: The blood compatibility of heparin loaded PLGA is modified significantly.