Preparation and performance of coronary stent PLGA coating using electrospray deposition(ESD)

HUANG Ying-ying,QI Min,YANG Da-zhi,WANG Chang-ming
DOI: https://doi.org/10.3321/j.issn:1001-9731.2006.03.022
2006-01-01
Abstract:The PLGA coatings on coronary stent surface were prepared using electrospray deposition(ESD) method.The morphologies of the deposited coatings were observed in term of OLYMPUS microscopy and atomic force microscope(AFM).To evaluate the binding force between coating and stent,the stent was expanded by balloon.Additionally,blood platelet adhesion and dynamic clotting time tests were carried out to evaluate the blood compatibility.These results indicate that the surface roughness of PLGA coating and 316L stainless steel are R_a=16.174nm,R_a=0.149nm respectively.There aren't avulsion and cock when the coatings stent were expanded.On the PLGA coatings,blood platelet adhesion is few,no distortion and no activation.The dynamic clotting time is prolonged until 50min.The blood compatibility of the PLGA coatings is excellent.