Determination of Trace Metal Impurities in Solar Grade Polysilicon by ICP-MS

Liu Zhijuan,Chu Lianqing,He Xiukun,Cao Quanxi
DOI: https://doi.org/10.16180/j.cnki.issn1007-7820.2010.08.018
2010-01-01
Abstract:This paper describes the method for the determination of trace Al、Fe、Ca、Mg、Cu、Zn、Cr、Ni and Mn in solar grade polysilicon by inductively coupled plasma mass spectrometry.The mass spectral interference and the matrix effect are discussed.Sc and Rh are used as internal standard to corrected the interference due to the effects of matrix and interface efficiently.Under the optimum conditions the detection limits for impurities were in the range of 0.030 g/L~0.500 g/L.The recoveries were within 90.0%~107.1% and relative standard deviation within 0.9%~3.4%.
What problem does this paper attempt to address?