Quantitative analysis of single aerosol particles using polycapillary X-ray optics

Tianxi Sun,Zhiguo Liu,Yude Li,Guangfu Wang,Guanghua Zhu,Xunliang Ding,Qing Xu,Hui Liu,Ping Luo,Qiuli Pan,Xiaoyan Lin,Yuepeng Teng
DOI: https://doi.org/10.1016/j.sab.2009.08.005
2009-01-01
Abstract:A laboratory micro X-ray fluorescence spectrometer based on polycapillary X-ray optics (PXRO) was used to carry out the quantitative X-ray fluorescence analysis of single aerosol particles with smaller size than that of focal spot of PXRO. The minimum detection limits measured with the thin-film reference standards were in the range from 13.3 to 0.7ngcm−2 when the operating current and voltage were 70mA and 35kV, respectively. In order to reduce the effects of the inhomogeneous distributions of the X-ray intensity in the focal spot of the PXRO on the analysis results, the sensitivities were corrected by using a Gaussian function for the quantitative analysis of single aerosol particles. The accuracy of the analysis of single standard solution drops was on average 25% depending on the element and concentration. The precision of the analysis was better than 5%.