Determination of Composition in PM_(10) Aerosols by High-Energy Polarized Energy-Dispersive X-ray Fluorescence Spectrometry
JI Ang,ZHENG Nan,WANG He-jin,XU Zi-you,LI Guo-hui
DOI: https://doi.org/10.3969/j.issn.0254-5357.2011.05.003
2011-01-01
Abstract:In order to coordinate with X-ray Diffractomer(XRD) research on analyzing the crystal phase of inhaled aerosol PM10 qualitatively and quantitatively,a method of using High-Energy Polarized Energy-Dispersive X-ray Fluorescence Spectrometry(HE-P-EDXRF) was developed to quantitatively measure the major and minor elements in aerosol PM10 collected on Whatman glass fiber filters.The study focused on the effect of the blank values of air filter membrane for the measurement compositions of aerosol PM10.The results indicate that the standards with glass fiber membrane filter should be used as carriers if the elemental area density of the blank glass fiber filter is more than 0.1 μg/cm3,and the nuclepore polycarbonate aerosol membrane filter should be used if the elemental area density is less than 0.1 μg/cm3.The measurement of heavy trace elements using He-P-EDXRF was also investigated,and extended the scope of the measuring elements to 62.The detection limits of Na,Mg,S,Y,Nd,Sm,Eu,Gd,Tb,Dy,Ho,Er,Tm,Yb and Lu were more than 0.1 μg/m2,the detection limits of Ca,Cr,Mn,Fe,Co,Ni,Cu,Zn,Br and Rb were less than 0.01 μg/m2,and the detection limits of Cl,Al,Si,P,K,Sc,Ti,V,Ge,As,Br,Se,Sr,Zr,Mo,Pd,Rh,In,Nb,Ag,Cd,Te,Sb,Sn,Ba,Cs,La,Ce,Pr,Au,Pt,W,Tl,Pb,Bi,Th and U were 0.1-0.01 μg/cm2.