Study of On-line Chemical Analysis of Aerosol Individual Particles with ICP-MS

SU Yong-yang,LI Zhi-ming,ZHOU Guo-qing,XU Jiang,ZHAI Li-hua,HUANG Neng-bin,ZENG Shi,ZHU Feng-rong
DOI: https://doi.org/10.3969/j.issn.1004-4957.2009.04.013
2009-01-01
Abstract:A rapid on-line chemical analysis of individual particles with ICP-MS was presented in this paper.Nano-size terbium oxide particles were generated and the number concentrations of Tb particles were measured by ICP-MS and condensation particle counter(CPC) method simultaneously.The number counting efficiency of ICP-MS was calculated by comparing the signal pulses concentration from ICP-MS with those from CPC.Monodisperse terbium nitrate particles with diameters from submicron to micron were generated with a commercial vibrating orifice aerosol generator by using argon as carrier gas and introduced into ICP-MS directly for on-line analysis.The results showed that nanoparticle number counting efficiency of ICP-MS was approximate unity.Response of ICP-MS for Tb particles with diameters ranging from 0.4 to 3 μm was linear with the mass of Tb particles,indicating that the efficiency was independent of the mass of the particle within this size range.Preliminary study showed that ICP-MS could be used not only for continuous aerosols analysis but also for individual particles analysis with rate of 10 particles per second.The ICP-MS method has a potential for on-line chemical analysis of aerosol particles containing refractory metal and transuranium element.
What problem does this paper attempt to address?