The applied study of virtual-reality technology for accurate placement of C2 pedicle screws

王亚明,田增民,杨军,尹丰,郑奎宏,王洪伟,卢旺盛
DOI: https://doi.org/10.3969/j.issn.1009-122X.2007.09.010
2007-01-01
Abstract:Objective To investigate a new method for axial pedicle screw placement based on virtual-reality technology. Methods A modified four-column stereotactic frame was used for immobilizing cranio-cervical specimen in 8 adult cadavers with screws to make the head-neck-shoulder form a rigid integrated body, and to maintain the stability of the spatial position. The 3D localization information of the axis was acquired by thin-slice CT scan. AeroTech surgical planning software was used for three-dimensional modeling. A safe, visualized, and individual virtual inserting path for screw placement was designed, and its security was validated through interative virtual maneuvering. Then the guiding-arch was used to hold the accurate direction for placing the guidewire. CT scan was rechecked to assess the accuracy of the screw placement. Results Among the 16 axial pedicle screw placements, the inserting divergence of the guidewire (breaching of the transverse foramen) occurred in one. The failure rate was 6.25%. Conclusion Three-dimensional volume reconstruction of the target axial pedicle, visualized design and virtual demonstration of inserting approach made surgical procedure simplified, intuitionistic and rigorous, not requiring linear and goniometrical parameters. The screw placement techniques are of good perspective of clinical application with the help of stability provided by guide-holding equipment.