Experimental study of cervical pedicle three-dimensional locator techniques

毛广平,赵建宁,王与荣,陈跃先,黎介寿,邱大学
DOI: https://doi.org/10.3969/j.issn.1005-8478.2005.18.015
2005-01-01
Abstract:[Objective]To develop a cervical pedicle three-dimensional locator for safely guiding cervicle screw placement with high accuracy and simplified manipulation.[Method]The screws were inserted into 90 cadaver cervical pedicles, under guidance of three-dimensional locator made by stainless steel ( group A ), where another 90 screws were under the Abumi manipulation ( group B ).[Result]In group A, 90% of the screws were placed accurately within the pedicles, 10% of the screws caused a pedicle cortex break; while in group B, the corresponding figures were 55.6% and 44.4%.[Conclusion] Cervical pedicle three-dimensional locator shows the advantaged of easy manipulation, high accuracy of screw placement and low cost, it is recommended for clinical application.