Evaluation on the influence of electrocardiograph modulated milliampere on image quality and exposure dosage of volume CT heart scan

ZHANG Sen,DU Xiang-ke,LI Jian-ying
DOI: https://doi.org/10.3760/j.issn:1005-1201.2006.09.018
2006-01-01
Abstract:Objective To find out whether the use of ECG modulated current(mA)will influence image quality and to decide whether the electrocardiograph(ECG)modulated mA will effectively reduce the exposure dosage.Methods The cardiac pulsating phantom was set at three speed levels,i.e.high, medium,and low speed so as to simulate different heart rates.The phantom was scanned with ECG modulated mA turned on and off,and the exposure dosage of each scan sequence was documented.The images were reeonstructed with reconstruction algorithm that matched the different levels of heart rate.CT values and their corresponding standard deviations at uniform areas on the images and the variation of the CT values at different locations were measured.The results from the two groups with and without ECG modulated mA were analyzed.Results Under the same level of heart rate,the exposure dosage was remarkably reduced when the EGG modulated mA was on than when it was off.Statistical analysis showed no significant difference(P>0.05)between the images from the two groups.Conclusion When scanning the heart with volume CT(VCT),the application of ECG modulated mA can effectively reduce the exposure dosage without sacrificing the image quality.