Determination of Azithromycin Concentration in Human Plasma by HPLC-ECD

PAN Yibin,CHEN Xingyan,HU Guoxin
DOI: https://doi.org/10.19577/j.cnki.issn10074406.2010.02.008
2010-01-01
Abstract:AIM To develop a HPLC-ECD method for the determination of azithromycin concentration in human plasma.METHODS A ZORBAX XDB-C18 column(Agilent 150 mm×4.6 mm,5 μm)was used with a mobile phase consisting of phosphate buffer(pH 7.0,50 mmol·L-1)-acetonitrile(70∶30,V/V).The flow rate was 1.0 mL·min-1.The column temperature was 45.0℃ and the electric potential of ECD was 650 mV.Clarithromycin was used as the internal standard.RESULTS The linear range of the plasma concentration of azithromycin was 10-100 0 μg·L-1 and the limit of quantification was 10 μg·L-1.The regression equation was ρ=6.132 8 Ai/As-0.019 5,r=0.999 9.The relative range of recovery was 85%-115% and the absolute recovery of component added to plasma samples was more than 75%.Both the inter-day RSD and the intra-day RSD were less than 10%.CONCLUSION The method has high sensitivity and good selectivity,and it is suitable for therapeatic drug monitoring and the pharmacokinetic research of azithromycin in human plasma.