Sensitive liquid chromatography/mass spectrometry assay for the quantification of azithromycin in human plasma.

Fei Liu,Yu Xu,Jinchang Huang,Shu Gao,Qingxiang Guo
DOI: https://doi.org/10.1002/bmc.883
2007-01-01
Biomedical Chromatography
Abstract:A simple, rapid, sensitive and specific liquid chromatography/electrospray ionization mass spectrometry method was developed and validated to quantify azithromycin in human plasma, using erythromycin as the internal standard (IS). A simple sample preparation method of protein precipitation with methanol was employed. Methanol, acetonitrile and water (12:30:58, v/v/ v) were used as the isocratic mobile phase, with 0.1% formic acid and 0.1% ammonium acetate in water. Selected ion monitoring was specific for azithromycin and erythromycin. The assay was linear over the concentration range 4.69-600 ng/mL. The correlation coefficients for the calibration curves ranged from 0.9994 to 0.9998. The intra- and inter-day precisions, calculated from quality control samples, were less than 8.24%. The method was employed in a pharmacokinetic study after oral administration of 500 mg azithromycin dispersible tablet to 20 healthy volunteers. Copyright (c) 2007 John Wiley & Sons, Ltd.
What problem does this paper attempt to address?