Development of a Sandwich Chemiluminescence Immunoassay for the Detection of Hyaluronic Acid from Human Serum

于峰,李瑞娟,李鹏聪,张鑫,孙旭东,刘宾
DOI: https://doi.org/10.3969/cmba.j.issn.1673-713x.2012.05.005
2012-01-01
Abstract:Objective To develop a chemiluminescence immunoassay (CLIA) for the detection of hyaluronic acid (HA) from human sera. Methods The quantitive CLIA was established based on double-antibody sandwich technology. Hyaluronic acid binding protein (HABP) was coated on microtiter plates and labeled with horseradish perxidase (HRP) respectively. Then the methodological evaluation and clinical application study about the method was assessed. Results The sensitivity of the CLIA method is 0.66μg/L, and linear range is 1.8~1000μg/L. Coefficients of variation (CV) for within-run and between-run assays is lower than 7.05%and 8.0%, respectively. The addition recovery and dilution recovery rate is 103.1%and 106.8%, respectively. The cross reactivity with Laminin, type IV collagen and type III procollagen is 0.19%, 0.16%and 0.17%, respectively. There is little effect of anticoagulants, bilirubin, cholesterol or triglyceride added in the serum samples on the assay. The conformance rate of 737 clinical serum samples is 98.64%as assayed by the method in Fushun infectious disease hospital. The correlation of HA detection assay in 197 serum samples between the CLIA method and HA Test Kit (Corgenix, USA) is y =0.855x+26.874 (r=0.9877). Conclusions The developed CLIA for the detection of HA from human sera is a sensitive and specific method with broad linear range, high accuracy and reproducibility. It can be applied in clinic for the detection of liver fibrosis.
What problem does this paper attempt to address?