Development of a Method of Chemiluminescence Immunoassay for CA15-3 with Native Produced Apparatus and Kits in This Laboratory

Huang Ying,Sun Xudong,Yang Xiaolin
DOI: https://doi.org/10.3969/j.issn.1008-9810.2006.05.051
2006-01-01
Abstract:ObjectiveTo establish a method of chemiluminescence immunoassay(CLIA) developed in this laboratory for serum CA15-3 determination with native produced apparatus and kits.Methods Serum CA15-3 contents were determined with CLIA,using native made apparatus and kits developed in our laboratory.Altogether 1232 specimens were tested(478 from patients and 754 from controls);the results were compared with those obtained with Roche-Elecsys 2010.Results The sensitiveity of the assay was 0.034KU/L.Linear range was 1.2~160KU/L.There were no cross-reactions with CA125 and CEA.Adding necessary amount of anticoagulants(sodium citrate,heparin,EDTA-Na_2) to the specimen had no effect on the assay results.Additional or dilutional recovery rate was 90%~110%.Both the inter and intra-CV were less than 10.0%.Normal range of CA15-3 concentration with this method was(0~23.08)KU/L(95% confidence limit).Results of this assay had a coincidence rate of 96.65% to those with the gold-standard Elecsys 2010 ECLIA.Conclusion The newly developed method of CLIA was sensitive,specific stable,accurate with good reproducibility.The imported apparatus and kits could be adequately replaced.
What problem does this paper attempt to address?