FLOW-INJECTION ENHANCEMENT CHEMILUMINESCENCE DETERMINATION OF HUMAN SERUM ALBUMIN BASED ON FCLA-1O2 REACTION

shingming luo,qun chen,da xing,yanchun wei,wei xu
DOI: https://doi.org/10.1142/9789812832344_0014
2008-01-01
Abstract:A flow-injection chemiluminescence method using reactive oxygen species (ROS) specific CL probe-FCLA for low level human serum albumin (HSA) detection is proposed in this study. The technique is based on the phenomena that low level HSA can effectively enhance CL emission of FCLA-O-1(2), system. In comparison to the CL quenching technique used by others, the proposed method significantly improves the detection sensitivity (ca. 100 folds higher). After optimizing the current apparatus, we have achieved a reliable detection range of HSA between 1 x 10(-10)-1 x 10(-8) mol L-1. The limit of detection is 4.5 x 10(-11) mol L-1 (S/N=3) for HSA and the R.S.D. is 3.08% (n=27) for 6 x 10(-9) mol L-1 HSA. The results are in excellent agreement with that from established clinical technique.
What problem does this paper attempt to address?