Electrochemiluminescence Biosensor for Hyaluronidase Activity Detection and Inhibitor Assay Based on the Electrostatic Interaction Between Hyaluronic Acid and Ru(bpy)32+

Zhixin Li,Huixing Chen,Zesheng Zhuo,Da Huang,Fang Luo,Li Chen,Jian Wang,Longhua Gun,Bin Qiu,Zhenyu Lin
DOI: https://doi.org/10.1016/j.snb.2018.08.069
2018-01-01
Abstract:The development of simple and sensitive method for hyaluronidase (HAase) detection is significant for the clinical diagnosis and treatment of cancer as HAase is a potential tumor marker. In this study, a simple but sensitive electrochemiluminescence (ECL) biosensor has been designed for HAase detection based on the electrostatic interaction of anionic hyaluronic acid (HA) and cationic luminous reagent Ru(bpy)(3)(2+). Ru(bpy)(3)(2+) is adsorbed on HA through electrostatic interaction to form HA-Ru(bpy)(3)(2+) complex. The Ru(bpy)(3)(2+) in the HA-Ru(bpy)(3)(2+) complex cannot be filtered out because of its large molecular weight. In the presence of HAase, the HA in HA-Ru(bpy)(3)(2+) can be cleaved into fragments by enzymatic hydrolysis. As a result, the Ru(bpy)(3)(2+) attached on the HA fragments can be separated from the HA-Ru(bpy)(3)(2+) complex by centrifugation easily. The resulting ultrafiltrate containing Ru(bpy)(3)(2+) can be used to characterize the concentration of HAase in the sample. The ECL intensity of Ru(bpy)(3)(2+) has a linear relationship with the concentration of HAase in the range of 2.0-40 U/mL with a detection limit of 0.33 U/mL. The proposed ECL system has been applied to detect HAase in urine samples with satisfied results and HAase inhibitor with high efficiency.
What problem does this paper attempt to address?