Coincidence time resolution of 50 ps FWHM using a pair of multi-anode MCP-PMTs with Cherenkov radiator window

Lingyue Chen,Lishuang Ma,Guorui Huang,Zhehao Hua,Muchun Jin,Zhen Jin,Sen Qian,Ling Ren,Shuguang Si,Jianning Sun,Qi Wu,Xingchao Wang,Yifang Wang,Zhi Wang,Ning Wang,Kai Wu,Haoda Zhang
DOI: https://doi.org/10.1016/j.nima.2024.169173
2024-02-17
Abstract:The use of time-of-flight (TOF) technology in positron emission tomography (PET) is expected to enhance the performance of clinical tasks such as lesion detection and disease staging. One critical requirement for TOF-PET is achieving lower Coincidence Time Resolution (CTR). In this manuscript, we developed a kind of multi-anode micro channel plate photomultiplier tube (MCP-PMT) specifically designed for TOF-PET applications. The direct use of Cherenkov radiator as the optical window eliminate the optical interface between the radiator and the detector, and the multi-anode structure enables a great positional resolution. Using the fast analog-to-digital converters (FADCs) technology to record the entire waveforms of the signals, the CTR measured by a pair of MCP-PMTs was analyzed offline, and the best CTR of 40.9±12.9 ps Full Width at Half Maximum (FWHM), which is equivalent to a position resolution of 6.1±1.9mm line of response (LOR) was obtained.
physics, particles & fields, nuclear,nuclear science & technology,instruments & instrumentation