DESIGN OF A NEW DEPTH ENCODING DETECTOR FOR PET

Huafeng Liu,Chao Bao,Mitsuo Watanabe,Takaji Yamashita
2000-01-01
Abstract:The parallax or depth of interaction (DOI) problem affects the image resolution of positron emission tomography system. A two-stage depth-encoding detector is described and evaluated. The detector consists of two layers of LSO crystals and a position sensitive photomultiplier tube (PSP-MT). The upper LSO array is placed on the lower LSO array with a shift of half the element pitch in both X and Y directions. Since the centroid of the output light distribution from each crystal element locates at different position on the PS-PMT, the crystal element can be identified on the image plane derived with the centroid calculation. The performance of the detector was evaluated in terms of applicability of PET. The intrinsic spatial resolution of the detector is 1.15mm FWHM for the upper array, and that for the lower array is 1. 34mm FWHM at the gamma-ray normal incidence. The average energy resolution for the upper elements is 23. 6%, and that for the lower elements is 19. 7% at the 511keV full energy peak from 22Na source. Three-stage depth encoding detector or four-stage depth encoding detector can be realized by the similar technique.