Dielectric waveguide setup tested with a superconducting millimeter-wave Fabry-Pérot interferometer at milli-Kelvin temperatures

Jakob Lenschen,Rosalie Labbe,Nils Drotleff,Markus Fuhrmann,Jürgen Lisenfeld,Hannes Rotzinger,Alexey V. Ustinov
2024-11-23
Abstract:We proposed and tested a cryogenic setup comprising dielectric waveguides for mm-wave frequencies in the range of 75-110 GHz and temperatures down to 10 mK. The targeted applications are quantum technologies at millimeter-wave frequencies, which require measurements at low photon numbers and noise. We show that the high density polyethylene waveguides combine a frequency independent low photon loss with a very low heat conductance. Black high density polyethylene shows a higher attenuation, which is useful to block thermal photons in a cryogenic environment. The dielectric waveguides are thermally anchored and attenuated at several stages of the cryostat. They are individually protected by additional metallic shields to suppress mutual cross-talk and external interference. We have measured a Fabry-Pérot cavity with superconducting mirrors at 10 mK and found out that the quality of a signal transmitted through the dielectric waveguides is sufficient to measure resonator quality factors over one million at 110 GHz.
Applied Physics,Superconductivity,Instrumentation and Detectors
What problem does this paper attempt to address?
This paper attempts to solve the technical problems of millimeter - wave measurement under extremely low - temperature conditions. Specifically, although traditional hollow metal waveguides can transmit millimeter - wave signals well, they are inflexible and will generate unacceptable heat loads at the low - temperature stage, which limits their applications in quantum technology, especially in scenarios requiring low photon numbers and low - noise measurements. In addition, the transmission of thermal radiation from the high - temperature stage will also cause the temperature at the low - temperature stage to rise, affecting the experimental conditions. For this reason, the paper proposes a new method using dielectric waveguides (DWGs) for millimeter - wave signal transmission from room temperature to the millikelvin temperature range. The advantages of this method include: 1. **Ultra - low thermal conductivity**: Dielectric waveguides have very low thermal conductivity and can effectively reduce the heat load. 2. **Good transmission performance**: Dielectric waveguides exhibit good transmission characteristics in the millimeter - wave frequency band. 3. **Mechanical flexibility**: Dielectric waveguides have good mechanical flexibility, which is convenient for installation and adjustment. 4. **Effective infrared radiation suppression**: Black dielectric waveguides can effectively block thermal photons and are suitable for low - temperature environments. Through these improvements, the paper shows how to achieve high - quality millimeter - wave measurements at extremely low temperatures. In particular, by testing a superconducting Fabry - Perot interferometer, the feasibility of this system is verified. The experimental results show that the system can measure a resonator with a quality factor exceeding one million at a frequency of 110 GHz, demonstrating its potential application value in the field of quantum technology.