High-count-rate Particle Tracking in Proton and Carbon Radiotherapy with Timepix2 Operated in Ultra-Short Acquisition Time

C. Oancea,A.Resch,S. Barna,G. Magrin,L. Grevillot,D. Hladik,L. Marek,J. Jakubek,C. Granja
2024-09-26
Abstract:This work investigates the operational acquisition time limits of Timepix3 and Timepix2 detectors operated in frame mode for high-count rate of high deposited energy transfer particles. Measurements were performed using alpha particles from a 241Am laboratory source and proton and carbon ion beams from a synchrotron accelerator. The particle count rate upper limit is determined by overlapping per-pixel particle signals, identifiable by the hits per pixel counter > 2, indicating the need to decrease acquisition time. On the other hand, the lower limit is the time required to collect the particle deposited charge while maintaining spectral properties. Different acquisition times were evaluated for an AdvaPIX Timepix3 detector (500 um Silicon sensor) with standard per-pixel DAC settings and a Minipix Timepix2 detector (300 um Silicon sensor) with standard and customized settings the pulse shaping parameter and threshold. For AdvaPIX Timepix3, spectra remained accurate down to 100 us frame acquisition time; at 10 us, loss of collected charge occurred, suggesting either avoiding this acquisition time or applying a correction. Timepix2 allowed acquisition times down to 100 ns for single particle track measurements even for high energy loss, enabled by a new Timepix2 feature delaying shutter closure until full particle charge collection. This work represents the first measurement utilizing Timepix-chips pixel detectors in an accelerator beam of clinical energy and intensity without the need to decrease the beam current. This is made possible by exploiting the short shutter feature in Timepix2 and a customized per-pixel energy calibration of the Timepix2 detector with a larger discharging signal value which allowed for a shorter time-over-threshold (ToT) signal. These customized settings extend the operation of the pixel detectors to higher event rates up to 10^9 particles/cm^2/s.
Instrumentation and Detectors,Accelerator Physics,Medical Physics
What problem does this paper attempt to address?
This paper aims to explore the operating limits of the Timepix2 detector under high count - rate conditions (such as in proton and carbon - ion radiotherapy), especially its performance under ultrashort acquisition times. Specifically, the researchers experimentally evaluated the effects of different acquisition times on the Timepix3 and Timepix2 detectors, especially for single - particle tracking and energy - deposition measurement of high - energy particles. The key points of the study are as follows: 1. **Determining the operating limits under high count - rate conditions**: The researchers conducted experiments using an α - particle source (241Am), proton beams, and carbon - ion beams to determine the minimum acquisition time limit of the Timepix2 detector under high count - rate conditions. This includes evaluating the overlap of particle signals at different acquisition times and the shortest acquisition time for maintaining spectral characteristics. 2. **Evaluating the effects of customized calibration parameters**: To improve the performance of Timepix2 under high count - rate conditions, the researchers customized the calibration of the detector, adjusting the pulse - shaping parameter (Ikrum) and the threshold (THL). These adjustments help to reduce the problem of incomplete charge collection, thereby improving the accuracy of single - particle detection. 3. **Verifying the application ability of Timepix2 at clinical intensities**: For the first time, the study demonstrated the application of the Timepix2 detector in proton and carbon - ion beams at clinical intensities without reducing the beam current. This is due to the new features of Timepix2, such as delaying the shutter closing until the particle charge collection is completed, and customized per - pixel energy calibration. 4. **Comparing the performance of different detectors**: The study also compared the performance differences between Timepix3 and Timepix2 at different acquisition times, especially at low acquisition times. The results show that Timepix2 can still maintain high spectral accuracy and single - particle detection ability at extremely short acquisition times (such as 100 nanoseconds), while Timepix3 has the problem of incomplete charge collection at acquisition times below 10 microseconds. In summary, the main objective of this paper is to experimentally verify the performance of the Timepix2 detector under high count - rate conditions and propose optimization schemes to achieve more efficient and accurate single - particle detection and energy - deposition measurement, especially in clinical proton and carbon - ion therapies.