Medipix3 proton and carbon ion measurements across full energy ranges and at clinical flux rates in MedAustron IR1

N.J.S. Bal,C.S. Schmitzer,A. De Franco,S. Enke
DOI: https://doi.org/10.1088/1748-0221/16/12/C12002
2021-12-04
Journal of Instrumentation
Abstract:The Medipix3, a hybrid pixel detector with a silicon sensor, has been evaluated as a beam instrumentation device with proton and carbon ion measurements in the non-clinical research room (IR1) of MedAustron Ion Therapy Center. Protons energies are varied from 62.4 to 800 MeV with 10 4 to 10 8 protons per second impinging on the detector surface. For carbon ions, energies are varied from 120 to 400 MeV/amu with 10 7 to 10 8 carbon ions per second. Measurements include simultaneous high resolution, beam profile and beam intensity with various beam parameters at up to 1000 FPS (frames per second), count rate linearity and an assessment of radiation damage after the measurement day using an x-ray tube to provide a homogeneous radiation measurement. The count rate linearity is found to be linear within the uncertainties (dominated by accelerator related sources due to special setup) for the measurements without degraders. Various frequency comp...