Evaluation of Timepix Silicon Detector for the Detection of 18F Positrons

Q. Wang,J. Tous,Z. Liu,S. Ziegler,K. Shi
DOI: https://doi.org/10.1088/1748-0221/9/05/c05067
2014-01-01
Journal of Instrumentation
Abstract:Timepix is an evolving energy and position sensitive pixel detector. It consists of a silicon detector (sensitive layer 300 mu m thick) bump-bonded to the Timepix readout chip developed by the Medipix2 collaboration. This study aims to test the feasibility of using the acquired energy and position signals from Timepix for positron imaging.The signals of the commonly used fluorine-18 PET (positron emission tomography) tracer [F-18]FDG were measured using Timepix operated both in single particle counting (Medipix) and in time over threshold (TOT) modes. The spatial resolution (SR) was measured using the absorber edge method (AEM) and was calculated from the over-sampled line spread function. The track of a positron in the Timepix detector was characterized as a cluster and the energy weighted centroid of each cluster was considered as readout for the position of the positron incidence. The measurement results were compared with theoretical predictions using Monte-Carlo simulations. In addition, imaging of a tissue slice of a mouse heart was analysed with reference to standard phosphor plate imaging.Our results show that the SR was improved from 177.1+/-4.1 mu m (centroid without energy weighting) to 155.5+/-3.1 mu m mu m (centroid with energy weighting). About 12% enhancement of SR was achieved with energy information in TOT mode. The sensitivity of Timepix was 0.35 cps/Bq based on the measurements. The measuring background and the ratio between detected positrons and gamma rays were also evaluated and were found to be consistent with theoretical predictions. A small enhancement of image quality was also achieved by applying energy information to the data of the measured tissue sample.Our results show that the inclusion of energy information could slightly enhance the positron measurement compared to without energy information and the Timepix provides a high SR and sensitivity for positron detection. Thus, Timepix is a potentially effective tool for 2D positron imaging.
What problem does this paper attempt to address?