Optimization of Time- and Code-Division-Multiplexed Readout for Athena X-IFU
W B Doriese,S R Bandler,S Chaudhuri,C S Dawson,E V Denison,S M Duff,M Durkin,C T FitzGerald,J W Fowler,J D Gard,G C Hilton,K D Irwin,Y I Joe,K M Morgan,G C O'Neil,C G Pappas,C D Reintsema,D A Rudman,S J Smith,R W Stevens,D S Swetz,P Szypryt,J N Ullom,L R Vale,J C Weber,B A Young
DOI: https://doi.org/10.1109/TASC.2019.2905577
2019-03-18
Abstract:Readout of a large, spacecraft-based array of superconducting transition-edge sensors (TESs) requires careful management of the layout area and power dissipation of the cryogenic-circuit components. We present three optimizations of our time- (TDM) and code-division-multiplexing (CDM) systems for the X-ray Integral Field Unit (X-IFU), a several-thousand-pixel-TES array for the planned Athena-satellite mission. The first optimization is a new readout scheme that is a hybrid of CDM and TDM. This C/TDM architecture balances CDM's noise advantage with TDM's layout compactness. The second is a redesign of a component: the shunt resistor that provides a dc-voltage bias to the TESs. A new layout and a thicker Pd-Au resistive layer combine to reduce this resistor's area by more than a factor of 5. Third, we have studied the power dissipated by the first-stage SQUIDs (superconducting quantum-interference devices) and the readout noise versus the critical current of the first-stage SqUIDs. As a result, the X-IFU TDM and C/TDM SQUIDs will have a specified junction critical current of 5 μA. Based on these design optimizations and TDM experiments described by Durkin, et al. (these proceedings), TDM meets all requirements to be X-IFU's backup-readout option. Hybrid C/TDM is another viable option that could save spacecraft resources.