AFM probe with the U-shaped cross-sectional cantilever for measuring the ultra-low coefficient of friction of 10−6

Yushan Chen,Liang Jiang,Linmao Qian
DOI: https://doi.org/10.1007/s40544-023-0821-z
IF: 4.924
2024-02-08
Friction
Abstract:Abstract Accurately measuring the coefficient of friction (COF) is the fundamental prerequisite of superlubricity research. This study aimed to reduce the COF measurement resolution Δ μ of atomic force microscopy (AFM). Based on the theoretical model, a distinctive strategy was adopted to reduce Δ μ by optimizing the cantilever’s cross-section of the AFM probe, inspired by civil engineering. Δ μ can be reduced by decreasing the width of the horizontal side w R and the wall thickness t and increasing the width of the vertical side w H . Moreover, the I-shape demonstrates the highest reduction in Δ μ , followed by the U-shape. Considering the processability, the AFM probe with the U-shaped cross-sectional cantilever was investigated further, and the dimensions are 35 μm w R , 3.5 μm w H , 0.5 μm t , 50 μm l (cantilever length), and 23 μm h tip (tip height). The finite element analysis results confirm its reliability. After being fabricated and calibrated, the AFM probe achieves the minimal Δ μ of 1.9×10 −6 under the maximum normal force so far. Additionally, the friction detection capability of the fabricated AFM probe improves by 78 times compared to the commercial tipless-force modulation mode (TL-FM) AFM probe with the conventional solid rectangular cross-sectional cantilever. This study provides a powerful tool for measuring 10 −6 COF.
engineering, mechanical
What problem does this paper attempt to address?