Time resolution and characteristic study of MWPC detectors with different Argon based gas

Rajendra Nath Patra,R. N. Singaraju,S. Biswas,T. K. Nayak,Y. P. Viyogi
DOI: https://doi.org/10.1007/978-3-319-73171-1_119
2018-10-07
Abstract:A prototype of Multi-Wire Proportional Chambers (MWPC) has been fabricated for the study of its various characteristics. The detector contains gold-coated tungsten wires (20 $\mu m$ diameter) on the anode frame, with a pitch of 2.8 mm. The gap between the anode and the cathode is 3 mm and the gap between anode and read-out is also 3 mm. Detailed study of MWPC in terms of gain, energy and timing resolution and efficiency measurements have been performed. The detector has been operated using Ar/CO$_{2}$ gas mixtures with 70:30 and 90:10 ratio. Energy spectrum of $^{55}$Fe X-ray source is obtained for the detector. The gain and energy resolution of the detector were calculated using X-ray spectrum. Time resolution is obtained $\sim$10 ns.
Instrumentation and Detectors,High Energy Physics - Experiment,Nuclear Experiment