Influence of the Working Gas Properties on the Anode Wire Modulation Effect of MWPC

Xiaohu Wang
DOI: https://doi.org/10.48550/arXiv.1501.01683
2015-01-07
Instrumentation and Detectors
Abstract:For MWPC used for X-ray position detection, simulation study of the anode wire modulation effect of the detector was carried out with Garfield program. Different gas mixtures were used as the working gas in the simulation, so as to obtain the influence of the X-ray cross section and electron diffusion coefficient of the working gases on the anode wire modulation effect of MWPC which has a anode wire spacing of 2mm. Results show that, though working gas with higher X-ray cross section implies a larger average drift distance of ionized electrons, using of such gas mixtures is of little use to improve the anode wire modulation effect of MWPC. And the transverse electron diffusion coefficient is the determining factor that affects the extent of the anode wire modulation effect of the detector.
What problem does this paper attempt to address?