Long term characterization of voltage references

Hubert Halloin,Pierre Prat,Julien Brossard
DOI: https://doi.org/10.48550/arXiv.1312.5101
2013-12-18
Abstract:We report here the characterization (temperature coefficients and noise level) of selected voltage references in the frequency range from 10^(-5) to 10 Hz. The goal of this work is to update previous studies, with a characterization at lower frequencies, and find voltage references that may be suitable for the space-based interferometry mission eLISA. The requirements of relative output stability of 1 ppm/$\sqrt{\text{Hz}}$ down to 0.1 mHz were not met by any of the tested devices, but 4 references approaches the objective : the AD587UQ, the MAX6126AASA50, the LT1021-BCN8-5 and the LT6655BHM. While the first three were already identified as potential devices in previous studies, the later is a new promising candidate using a different technology (bandgap).
Instrumentation and Detectors,Instrumentation and Methods for Astrophysics