Dark Current Measurements on a Superconducting Cavity using a Cryogenic Current Comparator

R. Geithner,R. Neubert,W. Vodel,P. Seidel,K. Knaack,S. Vilcins,K. Wittenburg,O. Kugeler,J. Knobloch
DOI: https://doi.org/10.1063/1.3527063
2010-07-29
Abstract:This paper presents non-destructive dark current measurements of superconducting TESLA cavities for superconducting LINACs. The measurements were carried out in a highly disturbed accelerator environment using a low temperature superconducting d.c. SQUID based Cryogenic Current Comparator (CCC). The overall current sensitivity under these rough conditions was measured to be 0.2 nA/sqrt(Hz) which enables the detection of dark currents of 5 nA.
Instrumentation and Detectors