A flow sampling strategy for the analysis of oil samples without pre-treatment in a sequential injection analysis system

Paula C.A.G. Pinto,M. Lúcia M.F.S. Saraiva,José L.F.C. Lima
DOI: https://doi.org/10.1016/j.aca.2005.09.009
IF: 6.911
2006-01-01
Analytica Chimica Acta
Abstract:The present work describes a sequential injection analysis (SIA) system adapted to direct analysis of oil samples. In this way, the flow system was designed by incorporation, in the SIA system, of an injection valve that was responsible for the sample insertion. With the developed sampling strategy the sample pre-treatment outside the flow system is avoided and also the problems associated with viscous samples in flow systems.The developed SIA system was applied to the determination of iron (Fe(III)) in edible oil samples and was based on the formation of a red complex (λ=510nm) between Fe(III) and thiocyanate in organic medium. A mixture of methanol:chloroform (85:15) was used as carrier solution and possible refractive index associated with the spectrophotometric detection was avoided by introduction of a mixing chamber in the flow system. The presented methodology produced 4.2ml of effluent and consumed 150μl of sample and 0.95mg of thiocyanate per determination.Linear calibration plots were obtained for Fe(III) concentrations up to 25mgl−1 and the detection limit of the determination was 0.31mgl−1. The developed methodology exhibited a good precision, with an R.S.D.<3.5% (n=15) and a determination frequency of 20 determinations h−1. The results of the analysis were evaluated by recovery studies (96.5–104.5%) and by the analysis of two AOCS reference samples.
chemistry, analytical