Comparison of hemodialysis urea clearance using spent dialysate and Kt/Vurea equations

Priyanka Khatri,Andrew Davenport
DOI: https://doi.org/10.1111/aor.14848
2024-08-26
Artificial Organs
Abstract:We measured urea clearance using standard Kt/Vurea equations with calculation of total body waster using Watson V or bioimpedance analysis and compared the values to Kt/V urea calculated using solute‐solver Lite online program and urea clearance measured from collection of effluent dialysate. Our study demonstrated that the result from equations estimating urea clearance indirectly from pre‐ and post‐blood samples from haemo‐ and haemodiafiltration treatments highly correlated with direct measurements of dialysate urea clearance. Introduction Dialysis adequacy is traditionally calculated from pre‐ and post‐hemodialysis session serum urea concentrations and expressed as the urea reduction ratio, or Kt/Vurea. However, with increasing hemodiafiltration usage, we wished to determine whether there were any differences between standard Kt/Vurea equations and directly measured spent dialysate urea clearance. Methods Urea clearance was measured from collected effluent dialysate and compared with various other methods of Kt/Vurea calculation, including change in total body urea from measuring pre‐ and post‐total body water with bioimpedance and the Watson equation, by standard Kt/V equations, and online clearance measurements using effective ionic dialysance (OLC). Results We compared urea clearance in 41 patients, 56.1% male, mean age 69.3 ± 12.6 years with 87.8% treated by hemodiafiltration. Reduction in total body urea was greater when estimating changes in total body urea, compared to measured dialysate losses of 58.4% (48.5–67.6) vs 71.6% (62.1–78), p
engineering, biomedical,transplantation
What problem does this paper attempt to address?