Patient Cranial Angle and Intrafractional Stability in CyberKnife Robotic Radiosurgery: A Retrospective Analysis

Chen-Lin Kang,Ya-Yu Huang,Yi-Ren Chen,Shu-Huei Tsai,Chun-Chieh Huang,Yu-Jie Huang
DOI: https://doi.org/10.1177/15330338241235058
2024-03-11
Technology in Cancer Research & Treatment
Abstract:Technology in Cancer Research &Treatment, Volume 23, Issue , January-December 2024. Purpose: The aim of this study was to investigate whether variations in cranial angles and treatment accuracy during CyberKnife robotic radiosurgery necessitate adjustment of the margins of the planning target volume. Patients and Methods: Data from 66 patients receiving CyberKnife treatment for brain tumors were retrospectively analyzed. Patients were immobilized using a thermoplastic mask and headrest. The cranial angle was measured on planning CT and patients were divided into 2 groups: ≤10° (Group A) and >10° (Group B). Intrafractional motion was recorded using the CyberKnife tracking system over 50 min. Translational and rotational errors were compared between groups, and planning target volume margins were calculated. Results: In Group A, significant translational error differences were found along with the X-axis over time (P < .02). In Group B, significant differences occurred along with the Z-axis (P < .03). No significant rotational or 3-dimensional vector differences were found in either group. Group A had significantly lower Y-axis (P < .045) and roll axis (P 10°.
oncology
What problem does this paper attempt to address?