Feasibility of patient‐specific quality assurance (PSQA) for real‐time robotic stereotactic body radiotherapy (SBRT) based on tumor motion traces

Qianyi Xu,Jiajin Fan,Yevgeniy Vinogradskiy,Ashish K. Chawla,Gregory Kubicek,Haihua Yang,Kiet Huynh,Tamara LaCouture,Jimm Grimm,Wei Nie
DOI: https://doi.org/10.1002/acm2.14352
2024-05-04
Journal of Applied Clinical Medical Physics
Abstract:Purpose To design a patient specific quality assurance (PSQA) process for the CyberKnife Synchrony system and quantify its dosimetric accuracy using a motion platform driven by patient tumor traces with rotation. Methods The CyberKnife Synchrony system was evaluated using a motion platform (MODUSQA) and a SRS MapCHECK phantom. The platform was programed to move in the superior‐inferior (SI) direction based on tumor traces. The detector array housed by the StereoPhan was placed on the platform. Extra rotational angles in pitch (head down, 4.0° ± 0.15° or 1.2° ± 0.1°) were added to the moving phantom to examine robot capability of angle correction during delivery. A total of 15 Synchrony patients were performed SBRT PSQA on the moving phantom. All the results were benchmarked by the PSQA results based on static phantom. Results For smaller pitch angles, the mean gamma passing rates were 99.75% ± 0.87%, 98.63% ± 2.05%, and 93.11% ± 5.52%, for 3%/1 mm, 2%/1 mm, and 1%/1 mm, respectively. Large discrepancy in the passing rates was observed for different pitch angles due to limited angle correction by the robot. For larger pitch angles, the corresponding mean passing rates were dropped to 93.00% ± 10.91%, 88.05% ± 14.93%, and 80.38% ± 17.40%. When comparing with the static phantom, no significant statistic difference was observed for smaller pitch angles (p = 0.1 for 3%/1 mm), whereas a larger statistic difference was observed for larger pitch angles (p
radiology, nuclear medicine & medical imaging
What problem does this paper attempt to address?