Silica Raman scattering probe for absolute calibration of Thomson scattering spectrometers

S. Ghazaryan,M. Kaloyan,C. Niemann
DOI: https://doi.org/10.1088/1748-0221/16/08/P08045
2021-08-17
Journal of Instrumentation
Abstract:We have developed a solid state probe for an absolute irradiance calibration of the Thomson scattering system on the Large Plasma Device (LAPD), based on Raman scattering off silica. Measurements performed with a triple-grating spectrometer have investigated the intensities of a pulsed laser beam Raman scattered off crystalline and amorphous silica over a range of temperatures of relevance to the LAPD (299–498 K). The data were compared with Rayleigh and Raman scattering intensities in gaseous nitrogen. The measurements show that Raman scattering off quartz allows rapid and accurate alignment and calibration of Thomson scattering systems in plasma physics experiments that cannot be calibrated using conventional methods.
instruments & instrumentation
What problem does this paper attempt to address?