Development and evaluation of relative QA dosimeter for electron beam based on CsPbBr 3

Seung-woo Yang,Yo-han Shin,Jae-hoon Jung,Heung-Lae Cho,Sung-kwang Park
DOI: https://doi.org/10.1088/1748-0221/19/03/c03062
2024-03-30
Journal of Instrumentation
Abstract:Quality assurance (QA) of electron beams must be performed regularly to ensure accurate radiation therapy. Therefore, a QA dosimeter for electron beams is required for accurate electron beam QA measurement. In this study, the electron beam detection performance of CsPbBr 3 material was measured. And it was evaluated whether it could be applied as a QA dosimeter for electron beams. To evaluate feasibility, reproducibility, linearity, and PDD were measured, and evaluated at 6 MeV, 9 MeV, and 12 MeV energies of the linear accelerator. The reproducibility evaluation results showed relative standard deviations of 1.06% at 6 MeV, 1.39% at 9 MeV, and 1.49% at 12 MeV. Linearity result evaluation, the coefficient of determination R 2 was presented as 0.9996, 0.9994, and 0.9992 at energies of 6, 9, and 12 MeV, respectively. When PDD was measured using the manufactured dosimeter, D max points for each energy band of 6, 9, and 12 MeV were shown. The CsPbBr 3 relative dosimeter produced in this study shows detection performance that satisfies the evaluation criteria of reproducibility, linearity, and PDD. These results mean that the CsPbBr 3 dosimeter is feasible as a QA dosimeter dedicated to electron beams.
instruments & instrumentation
What problem does this paper attempt to address?