Key comparison BIPM.RI(I)-K8 of high dose-rate 192 Ir brachytherapy standards for reference air kerma rate of the LNE-LNHB and the BIPM

C Kessler,V Hernandez-Elvira
DOI: https://doi.org/10.1088/0026-1394/61/1a/06015
2024-10-10
Metrologia
Abstract:Main text An indirect comparison of the standards for reference air kerma rate for 192 Ir high dose rate (HDR) brachytherapy sources of the Laboratoire National Henri Becquerel (LNE-LNHB), France, and of the Bureau International des Poids et Mesures (BIPM) was carried out at the LNE-LNHB in May 2024. The comparison result, based on the calibration coefficients for a transfer standard and expressed as a ratio of the LNE-LNHB and the BIPM standards for reference air kerma rate, is 0.9990 with a combined standard uncertainty of 0.0071. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/. The final report has been peer-reviewed and approved for publication by the CCRI, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).