International comparison of measurements of neutron source emission rate (2016-2021) - CCRI(III)-K9.Cf.2016
N J Roberts,N A Horwood,Z Vykydal,H Park,J Kim,W W Pereira,E S da Fonseca,C Thiam,Zhang Hui,M S Dewey,H P Mumm,H Harano,T Matsumoto,A Masuda,S Manabe,J P Archambault,N N Moiseev,A V Didyk
DOI: https://doi.org/10.1088/0026-1394/61/1a/06001
2024-01-20
Metrologia
Abstract:Main text Section III (neutron measurements) of the Comité Consultatif des Rayonnements Ionisants, CCRI, conducted a comparison of primary measurements of the neutron emission rate of a 252 Cf radionuclide source. A single 252 Cf source was circulated to all participants between 2016 and 2020. Ten laboratories participated -CMI (Czech Republic), KRISS (Republic of Korea), IRD/LNMRI (Brazil), LNE-LNHB (France), NIM (China), NIST (USA), NMIJ (Japan), NPL (UK), NRC (Canada) and VNIIM (Russia) - with NPL making their measurements at the start and repeating them at the end of the exercise to verify the 250 Cf content of the source. Each laboratory reported the emission rate into 4π sr together with a detailed uncertainty budget. All participants used the manganese bath technique except NMIJ who used a relative method based on measurements with a 3 He detector in a graphite pile. VNIIM also made measurements using an associated particle technique. CMI, KRISS, LNE-LNHB, NIM, NPL and VNIIM also measured the anisotropy of the source although this did not formally form part of the comparison. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/. The final report has been peer-reviewed and approved for publication by the CCRI, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).
instruments & instrumentation,physics, applied