Abstract:Main text A new comparison of the standards for air kerma of the Bundesamt für Eich- und Vermessungswesen (BEV), Austria, and of the Bureau International des Poids et Mesures (BIPM) was carried out in the 137 Cs radiation beam using the radiation protection facility of the International Atomic Energy Agency (IAEA) in Seibersdof, Austria in April 2023. The comparison was done using the BEV and the BIPM primary standards and a transfer instrument belonging to the BEV. The direct comparison result, evaluated as a ratio of the BEV and the BIPM standards for air kerma, is 1.0043 with a combined standard uncertainty of 2.7 parts in 10 3 . The result for the indirect comparison is in agreement with the direct result at the level of 3 parts in 10 4 . The results are analysed and presented in terms of degrees of equivalence, suitable for entry in the BIPM key comparison database, to update the 2009 comparison result. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/. The final report has been peer-reviewed and approved for publication by the CCRI, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).