Key comparison BIPM.RI(I)-K8 of high dose-rate 192 Ir brachytherapy standards for reference air kerma rate of the PTB and the BIPM

C Kessler,R Behrens,A Kasper,F Grote
DOI: https://doi.org/10.1088/0026-1394/61/1a/06002
2024-01-26
Metrologia
Abstract:Main text An indirect comparison of the standards for reference air kerma rate for 192 Ir high dose rate (HDR) brachytherapy sources of the Physikalisch-Technische Bundesanstalt (PTB), Germany, and of the Bureau International des Poids et Mesures (BIPM) was carried out at the PTB in May 2023. The comparison result, based on the calibration coefficients for a transfer standard and expressed as a ratio of the PTB and the BIPM standards for reference air kerma rate, is 1.0022 with a combined standard uncertainty of 0.0101. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/. The final report has been peer-reviewed and approved for publication by the CCRI, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).