Liquid scintillation efficiencies, gamma-ray emission intensities, and half-life for Gd-153

Denis E Bergeron,Jeffrey T Cessna,Ryan P Fitzgerald,Gulakhshan Hamad,Lizbeth Laureano-Pérez,Leticia Pibida,Brian E Zimmerman,Denis E. Bergeron,Jeffrey T. Cessna,Ryan P. Fitzgerald,Brian E. Zimmerman
DOI: https://doi.org/10.1016/j.apradiso.2023.111108
IF: 1.787
2024-01-01
Applied Radiation and Isotopes
Abstract:Gadolinium-153 was standardized for activity by live-timed anticoincidence counting and an ampoule was submitted to the international reference system (SIR). Absolute emission intensities for the main γ rays were determined with calibrated high-purity germanium (HPGe) and lithium-drifted silicon (Si(Li)) detectors. A revised decay scheme is indicated, with no probability of direct electron capture to the <sup>153</sup>Eu ground state. Triple-to-double coincidence ratio (TDCR) efficiency curves indicate that the revised decay scheme is consistent with experiment. Half-life measurements agree with a previous NIST determination and show no sensitivity to chemical environment.
radiology, nuclear medicine & medical imaging,nuclear science & technology,chemistry, inorganic & nuclear