Response of gadolinium doped liquid scintillator to charged particles: measurement based on intrinsic U/Th contamination

Qiang Du,Shin-Ted Lin,Han-Tao He,Shu-Kui Liu,Chang-Jian Tang,Li Wang,Henry T. Wong,Hao-Yang Xing,Qian Yue,Jing-Jun Zhu
DOI: https://doi.org/10.1088/1748-0221/13/04/P04001
2018-03-06
Abstract:A measurement is reported for the response to charged particles of a liquid scintillator named EJ-335 doped with 0.5% gadolinium by weight. This liquid scintillator was used as the detection medium in a neutron detector. The measurement is based on the in-situ $\alpha$-particles from the intrinsic Uranium and Thorium contamination in the scintillator. The $\beta$-$\alpha$ and the $\alpha$-$\alpha$ cascade decays from the U/Th decay chains were used to select $\alpha$-particles. The contamination levels of U/Th were consequently measured to be $(5.54\pm0.15)\times 10^{-11}$ g/g, $(1.45\pm0.01)\times 10^{-10}$ g/g and $(1.07\pm0.01)\times 10^{-11}$ g/g for $^{232}$Th, $^{238}$U and $^{235}$U, respectively, assuming secular equilibrium. The stopping power of $\alpha$-particles in the liquid scintillator was simulated by the TRIM software. Then the Birks constant, $kB$, of the scintillator for $\alpha$-particles was determined to be $(7.28\pm0.23)$ mg/(cm$^{2}\cdot$MeV) by Birks' formulation. The response for protons is also presented assuming the $kB$ constant is the same as for $\alpha$-particles.
Instrumentation and Detectors,High Energy Physics - Experiment
What problem does this paper attempt to address?