Experimental research for the measurement of thermal neutron flux with gadolinium

Yong BI,Chang-Jian TANG,Qian YUE,Jian-Ping CHENG,Ke-Jun KANG,Yuan-Jing LI,Jin LI,Hao-Yang XING,Yi-Gang YANG
DOI: https://doi.org/10.3969/j.issn.0258-0934.2010.08.030
2010-01-01
Abstract:The measurement of thermal neutron flux is an important work in dark matter detection experiment. A new approach, using the prompt gamma rays from the (Gd, nth) reaction to decide the thermal neutron flux, is studied in the work. Two NaI(Tl) detectors are used for coincident detection of these prompt gamma rays from (Gd, nth) reaction, and a powerful shield system preventing gamma rays and scattered neutrons from NaI(Tl) detectors is designed. After cross-checked measurement of background and neutron events, the spectrum of gamma rays from the reaction of (Gd, nth) is obtained and analyzed, the counting rate contributed from this reaction is calculated and its maximum is 2.74 Hz. The feasibility of this approach is confirmed.