Performance evaluation of a 71Ga(n,γ)72Ga reaction-based epithermal neutron flux detector at an AB-BNCT device

Xingcai Guan,Huangxin Wu,Rui Bai,Guanghua Wu,Wenliang Yang,Wuliang Guo,Haixi Wang,Yongquan Wang,Junliang Du,Lu Zhang,Long Gu
DOI: https://doi.org/10.1016/j.apradiso.2024.111249
IF: 1.787
2024-05-01
Applied Radiation and Isotopes
Abstract:The <sup>71</sup>Ga(n,γ)<sup>72</sup>Ga reaction-based epithermal neutron flux detectors are novel instruments developed to measure the epithermal neutron flux of boron neutron capture therapy (BNCT) treatment beams. In this study, a spherical epithermal neutron flux detector using <sup>71</sup>Ga(n,γ)<sup>72</sup>Ga reaction was prototyped. The performance of the detector was experimentally evaluated at an accelerator-based BNCT (AB-BNCT) device developed by Lanzhou University, China. Based on the experimental results and related analysis, we demonstrated that the detector is a reliable tool for the quality assurance of BNCT treatment beams.
radiology, nuclear medicine & medical imaging,nuclear science & technology,chemistry, inorganic & nuclear