Real-time neutron beam monitor utilizing scintillator technology for boron neutron capture therapy

Sang June Jeon,Jaewon Choi,Kyu Young Lee,Jungyu Yi
DOI: https://doi.org/10.1016/j.radmeas.2024.107070
IF: 1.743
2024-02-03
Radiation Measurements
Abstract:In boron neutron capture therapy (BNCT), patient dose is contingent upon boron concentration and neutron irradiation. This study details the creation and implementation of a real-time beam-monitoring system using a Eu:LiCaAlF 6 (LiCAF) scintillation detector. This system precisely regulates neutron irradiation according to the treatment plan. A two-step calibration process was employed for enhanced reliability. The initial calibration used an Americium–Beryllium (AmBe) source to isolate neutron signals from a mix of gamma and neutron signals. The second calibration involved a polymethyl methacrylate (PMMA) moderator within the BNCT beam, calibrated against the gold-foil activation method. This yielded a thermal neutron reaction rate with an uncertainty of approximately 2.62 % (k = 2). For continuous monitoring, the system includes a detector with a polyethylene (PE) moderator and an internal LiCAF placed inside the beam port. This monitoring system calculates and displays the epithermal neutron flux by comparing the flux at the beam port to the thermal neutron flux from the moderator. The performance of the installed system was evaluated, noting an uncertainty in epithermal neutron conversion of 3.3 % (k = 2) and a linearity exceeding 0.99 (R 2 > 0.99). Operational from December 2022 to April 2023 in clinical trials, the system was subjected to regular quality assurance checks on an annual, monthly, and daily basis, showing no performance degradation or unusual operation. Future work will focus on collecting data to determine the lifespan of the LiCAF detector through continued monitoring.
nuclear science & technology
What problem does this paper attempt to address?