Characterization of a broad-energy germanium detector using an assembled collimation device at CJPL

Zhi Zeng,Yu-Hao Mi,Ming Zeng,Hao Ma,Q. Yue,Jianping Cheng,Junli Li,Rui Qiu,Hui Zhang
2016-01-01
Abstract:Broad-energy germanium (BEGe) detectors have a strong pulse-shaped discrimination power, which makes them competitive candidates for neutrinoless double beta decay (0{nu}b{eta}b{eta}) experiments using germanium detectors. To achieve good performances of BEGe detectors in experiments, characterization must be performed in advance. Here, the characterization of a commercial p-type BEGe detector is presented. Several point sources were used to investigate the energy resolution and linearity of the detector, which were excellent. A cylindrical volume source was used for the efficiency calibration, and an efficiency curve was generated. With an assembled device for source positioning, a collimated 133Ba point source was used to scan the detector to investigate the active volume. Point sources of 133Ba and 241Am were used to measure the dead-layer thicknesses, which are approximately 0.17 mm on the front and 1.15 mm on the side. The described characterization method will play an important role in possible 0{nu}b{eta}b{eta} experiments with BEGe detectors at CJPL in the future.