Identification of anomalous fast bulk events in a p -type point-contact germanium detector

Ren-Ming-Jie Li,Shu-Kui Liu,Shin-Ted Lin,Li-Tao Yang,Qian Yue,Chang-Hao Fang,Hai-Tao Jia,Xi Jiang,Qian-Yun Li,Yu Liu,Yu-Lu Yan,Kang-Kang Zhao,Lei Zhang,Chang-Jian Tang,Hao-Yang Xing,Jing-Jun Zhu
DOI: https://doi.org/10.1007/s41365-022-01041-x
2022-01-01
Nuclear Science and Techniques
Abstract:The ultralow detection threshold, ultralow intrinsic background, and excellent energy resolution of p -type point-contact germanium detectors are important for rare-event searches, in particular for the detection of direct dark matter interactions, coherent elastic neutrino-nucleus scattering, and neutrinoless double beta decay. Anomalous bulk events with an extremely fast rise time are observed in the CDEX-1B detector. We report a method of extracting fast bulk events from bulk events using a pulse shape simulation and reconstructed source experiment signature. Calibration data and the distribution of X-rays generated by intrinsic radioactivity verified that the fast bulk experienced a single hit near the passivation layer. The performance of this germanium detector indicates that it is capable of single-hit bulk spatial resolution and thus provides a background removal technique.