Enhancing LOD Determination in Gas Chromatography: Validating the Hubaux-Vos Method for Gas Concentration Measurement

Hyeong Rae Kim,Sunwoo Kim,Jinsang Jung,Hwan Lee,Keon Ho,Byeongmoon Kim,Sanghyub Oh
DOI: https://doi.org/10.1016/j.chroma.2024.464764
IF: 4.601
2024-02-01
Journal of Chromatography A
Abstract:The limit of detection (LOD) is a crucial measure in analytical methods, representing the smallest amount of a substance that can be distinguished from background noise. In the realm of gas chromatography (GC), however, determining LOD can be quite subjective, leading to significant variability among researchers. In this study, we validate the Hubaux-Vos method, an International Standards Organization(ISO)-approved approach for determining LOD in gas concentration measurements, using a GC equipped with a discharge ionization detector (DID) and a dynamic dilution system. We employ a gas mixture certified reference material (CRM) of CO, CH<sub>4</sub>, and CO<sub>2</sub> at various concentrations to generate calibration curves for each gas. Subsequently, we estimate the LODs for each gas using the Hubaux-Vos method. Surprisingly, our findings indicate a notable difference between the LODs calculated using the Hubaux-Vos method and those confirmed through experiments. This highlights the importance of critically examining the theoretical foundations of LOD determination. We strongly recommend researchers to scrutinize the principles guiding LOD determination. The method proposed in this study offers an effective way to rigorously validate theoretical approaches for estimating LODs in gas concentration measurements using GC.
chemistry, analytical,biochemical research methods
What problem does this paper attempt to address?