Characterization of real-time cine MR imaging distortion on 0.35 T MRgRT with concentric Cine imaging QA phantom

Shanti Marasini,Mike Cole,Austen Nathaniel Curcuru,Lara M Dyke,H Michael Gach,Rocco Flores,Taeho Kim
DOI: https://doi.org/10.1088/1361-6560/ad2d33
IF: 3.5
2024-02-28
Physics in Medicine and Biology
Abstract:Objective: Real-time MRgRT uses 2D-cine imaging for target tracking and motion evaluation. Rotation of gantry induced B0 off-resonance, resulting in image artifacts and imaging isocenter-shift precluding MR-guided arc therapy. This work aimed to examine the spatial accuracy including geometric distortion and isocenter shift in real-time during gantry rotation on a 0.35T MR-Linac using the concentric Cine imaging QA phantom and its associated image analysis software. Approach: The Cine imaging QA phantom consists of two concentric shells of low-T1 mineral oil and a central alignment structure. The phantom was scanned on three different MRI systems; 0.55T Siemens Free.Max, 1.5T Philips Ingenia, and 0.35T ViewRay MRIdian MR-Linac using 2D balanced steady-state free precession (bSSFP) imaging sequence. In addition, bSSFP cine MRI with the banding artifact correction was tested on 0.35T ViewRay MR-Linac. Images from the MR-Linac were acquired with the Linac gantry stationary and rotating from gantry 300° 0° and vice versa. Three orthogonal image planes were scanned excluding the 1.5T Philips Ingenia, where only the axial plane was scanned. The image analysis software calculated the distortion values as well as the isocenter position for each cine frame. Main results: The geometric distortion on MRIs and MR-Linac at gantry stationary are within 1 mm while the substantial geometric distortion of 2 mm and 2.9 mm were observed on 0.35 T MR-Linac while rotating the gantry clockwise (300 0°) and counterclockwise 0 300° respectively, The average imaging isocenter shift was 0.1 mm for both MRIs and the static gantry and isocenter shift of ≤1.5 mm observed during the gantry rotation. The imaging isocenter shift decreased by 1±0.2 mm clockwise and counterclockwise with B0 compensation. Significance: The concentric Cine imaging QA phantom and its associated software effectively demonstrate the image distortion on real-time cine imaging on regular MRIs and 0.35T MR-Linac.
engineering, biomedical,radiology, nuclear medicine & medical imaging
What problem does this paper attempt to address?