Evaluation of the impact of cardiac implantable electronic devices on cine MRI for real‐time adaptive cardiac radioablation on a 1.5 T MR‐linac

Osman Akdag,Stefano Mandija,Pim T. S. Borman,Paris Tzitzimpasis,Astrid L. H. M. W. van Lier,Rick Keesman,Bas W. Raaymakers,Martin F. Fast
DOI: https://doi.org/10.1002/mp.17438
IF: 4.506
2024-10-06
Medical Physics
Abstract:Background Stereotactic arrhythmia radioablation (STAR) is a novel treatment approach for refractory ventricular tachycardia (VT). The risk of treatment‐induced toxicity and geographic miss can be reduced with online MRI‐guidance on an MR‐linac. However, most VT patients carry cardiac implantable electronic devices (CIED), which compromise MR images. Purpose Robust MR‐linac imaging sequences are required for cardiac visualization and accurate motion monitoring in presence of a CIED during MRI‐guided STAR. We optimized two clinically available cine sequences for cardiorespiratory motion estimation in presence of a CIED on a 1.5 T MR‐linac. The image quality, motion estimation accuracy, and geometric fidelity using these cine sequences were evaluated. Methods Clinically available 2D balanced steady‐state free precession (bSSFP, voxel size = 3.0 × 3.0 × 10 mm3, Tscan = 96 ms, bandwidth (BW) = 1884 Hz/px) and T1 ‐spoiled gradient echo (T1 ‐GRE, voxel size = 4.0 × 4.0 × 10 mm3, Tscan = 97 ms, BW = 500 Hz/px) sequences were adjusted for real‐time cardiac visualization and cardiorespiratory motion estimation on a 1.5 T Unity MR‐linac (Elekta AB, Stockholm, Sweden), while complying with safety guidelines for MRI in presence of CIEDs (specific absorption rate