A near X-point charge exchange neutral spectroscopy (CENS) system for DIII-D

S. R. Haskey,C. Chrystal,A. Angulo,A. Bortolon,E. Wolfe,E. Linsenmayer,C. Marini,F. Scotti,K. Agustin
DOI: https://doi.org/10.1063/5.0219576
IF: 1.6
2024-09-01
Review of Scientific Instruments
Abstract:A 16 channel spectroscopy system has been installed on DIII-D to provide information about the energy distribution of the atomic neutrals using the Doppler shift and broadening of passive Balmer-α emission. The lines of sight are dominantly in the toroidal direction, with tangencies near the lowest point on closed magnetic flux surfaces moving from the lower divertor upward into the confined plasma. This allows the system to provide details of the neutrals as they undergo various atomic physics processes while traveling upward from the lower divertor. The spectrally resolved measurements provide several advantages that complement typical optical filter based measurements of hydrogenic spectral lines. These include more direct measurements of the neutrals that fuel deeper in the plasma and capturing neutral velocity distribution information via the Doppler broadened and shifted line. In addition, the spectral separation of higher energy “thermal” neutrals and bright cold emission from the scrape-off layer allows for a more straightforward calculation of the underlying neutral densities based on the emission because of the dependence of the thermal emission on confined plasma properties that are approximately flux functions and well measured.
instruments & instrumentation,physics, applied
What problem does this paper attempt to address?