Investigation of the 106 Ru Ophthalmic Plaque Dose Distribution in a Pseudotumor-Contained Eye-Equivalent Phantom by Silica Glass Beads Ru-106 ophthalmic plaque dosimetry by Silica glass beads

Samaneh Hashemi,Mahdi Kahani,Zahra Siavashpour,Saeed Karimi,Ramin Jaberi
DOI: https://doi.org/10.1016/j.radmeas.2024.107145
IF: 1.743
2024-05-05
Radiation Measurements
Abstract:Purpose To investigate the 106 Ru dose distribution in an intraocular pseudotumor by micro silica glass beads thermo-luminescence dosimeters (TLD). Materials and methods Dosimetry was conducted using Monte Carlo (MC) simulation as well as an experimental setup by employing an in-house eye equivalent phantom. A BEBIG CCA model of 106 Ru plaque was used. Bead TLDs were modeled using MCNPX code, choosing water and plexiglass as phantom mediums along with CCA plaque. The deposited energy and electron energy spectra were calculated using *f8 and f4 tallies, respectively. A dedicated eye-equivalent phantom was designed and fabricated by 3D-printing techniques to place glass beads. Results Absorbed dose at the central axis of the CCA plaque by beads TLDs was compared to manufacturer-reported data and simulation results. Maximum, minimum, and mean relative differences of 23%, 7%, and 15% with the manufacturer and 14%, 0%, and 1.6% with the simulation were obtained, respectively. Replacing glass bead material with water resulted in an average relative difference of 33.9% in absorbed energy per mass unit, corresponding to 21% and 44.8% at the first (0.5 mm depth) and last dosimetry points (9.5 mm depth), respectively. It has also been shown that glass beads can shadow each other, leading to a dose reduction of 15%. Conclusion A specific eye equivalent phantom with a pseudotumor was constructed, and silica glass beads were used for dosimetry. The results of the experiment and the MC simulation are in good accord. In the case of 106 Ru ophthalmic brachytherapy plaque dosimetry, replacing glass beads material with water led to a mean difference of about 34% as well as shadowing effects. This emphasizes how crucial it is to employ a reliable correction factor in clinical applications.
nuclear science & technology
What problem does this paper attempt to address?